BD LSR-II SORP Analyzer

The BD LSR-II SORP is an analyzer flow cytometer equipped with 4 lasers.

The cytometer allows detection of 1parameters simultaneously (15 colors plus forward scatter and side scatter).

The LSR-II SORP offers high acquisition speeds, full inter- and intra-laser compensation matrix, and multiple thresholds on all lasers. Digital pulse sampling allows the measurement of Height, Area and Width from all parameters simultaneously.

 

Features

  • Optical System: Patented BD OctagonTM and TrigonTM technology. A quartz cuvette is coupled to the emission lens by refractive index-matching optical gel for optimum collection efficiency.
  • laser excitation and filters (see LSR-II configuration in the related links):

 

Laser line/color
(Excitation)
Detector Emission Filter (nm) Example of available markers

  Violet (405nm)

 A 780 / 60 – 740 BV750, BV785
 B   705 / 70 – 685LP

BV711 , BV650

 C 610 / 20 – 595LP BV605
 D 525 / 50 – 505LP BV510, Zombie Aqua
  E 450 / 50 Pacific Blue, BV421, DAPI, Hoechst, CFP

Blue (488nm)

 A 710 / 50 – 685LP PerCP, PerCP-Cy5-5, 7-AAD
 B 525 / 50 – 505LP GFP, AF488, YFP
 C 488 / 10 SSC
 D FSC

Yellow/green (561nm)

 A 780 / 60 – 740LP PE-Cy7
 B 710 / 50 – 685LP

PE-Cy5.5

 C 660 / 20 – 635LP

PE-Cy5

 D 610 / 20 – 600LP

mCherry, PI

 E 582 / 15 -570LP

DsRed, PE

Red (640nm)

 A 780 / 60 – 740LP APC- cy7
 B 710 / 50 – 685LP Vio Bright R720, AF700
 C 670 / 30 AF647, APC

 

  • High Throughput System (HTS) –  enables analysis directly from 96 multi-well plate. The workstation is managed by FACS-Diva 6.1 software.

 

Additional available filters:

 

Mirror 450LP 5005LP 530LP 635LP 685LP 735LP
Band pass 405/20 525/50 530/30 585/42 610/20 670/14 675/20 695/40

 

Acquisition Rate: >20000 events per second

Fluorescence Sensitivity: Detection limit is <200 molecules of equivalent soluble fluorochrome (MESF).

Fluorescence Resolution: Coefficient of variation (CV) of FL-Area is <3%, determined by full G0/G1 peak for propidium iodide (PI)-stained chicken erythrocyte nuclei (CEN Quality Control Kit).

 

Fluorescence Linearity: Doublet/singlet ratio for CENs stained with PI or Hoechst is 1.95-2.05, detected with 488nm laser.

 

Forward and Side Scatter Sensitivity: Sensitivity enables the separation of fixed platelets from noise.

 

Signal Processing

  • Digital electronics with 10 MHz sampling frequency
  • 18-bit data acquisition for area signals
  • Inter- and intra-beam compensation on-line and post-acquisition off-line
  • Height, Width, and Area measurement available for all parameters
  • Time parameter available for kinetic studies (Calcium)
  • Ratio measurements between any two parameters from any laser

 

Software

  • BD FACSDiVaTM Software Version 6.1
  • All data are saved linearly
  • Complex hierarchical gating for multicolor applications
  • Automatic compensation of all parameters after acquisition
  • Complete set of statistics for all analysis

 

Applications

Analysis of cells, yeast, bacteria and beads

  • Multi-color analysis
  • HTS system
  • Calcium Flux
  • DNA analysis

 